Removal of infected pacemaker leads using an endoscopic minimally invasive cardiac surgical approach: a case report

Background

Cardiac implantable electronic device (CIED) leads occasionally need to be removed when infected or in the case of lead failure. Transvenous lead extraction (TLE) using an excimer laser or mechanical sheath is often performed; however, when that approach is difficult, surgical removal through a sternotomy is required. The risk of deep sternal wound infection (DSWI) is especially high in cases in which the infected site is close to the sternum. We report a case in which surgical removal of CIED leads via a thoracoscopic approach without a sternotomy was effective.

Case presentation

The patient was a 52-year-old man with complete atrioventricular block. A skin fistula with purulent drainage developed above the pacemaker (PM) generator in the left anterior thoracic region and a device infection was suspected.

He had a PM implanted via lower sternotomy and the generator was placed in his abdomen at the age of 11. Twenty-eight years ago (at age 23), a PM was reimplanted transvenously from the right anterior thoracic region. A PM reimplantation was made to the left anterior thoracic region due to skin thinning at age 39, and 2 leads was remained. Thus, a total of four transvenous leads were implanted (one right atrial (RA)/ right ventricular (RV) lead from each side) (Fig. 1a).

Fig. 1
figure 1

Perioperative chest radiography. A toral of four leads were placed transvenously and an epicardial lead was left in the diaphragmatic plane (a). Two RV leads was remained after transvenous lead extraction (b). Only a part of the RV lead remained and a new epicardial lead was placed (c)

A tissue culture of skin fistula was positive for methicillin-sensitive Staphylococcus aureus (MSSA). An attempt was made to remove the infected device and leads via a transvenous approach using both an excimer laser and a mechanical sheath. The generator and two RA leads were extracted but the two RV leads couldn’t be extracted (Fig. 1b). He was therefore referred to our department for surgical removal. The open wound in the left anterior thorax had a pocket, the inner edge of which reached the anterior surface of the sternum (Fig. 2). A surgical approach with a sternotomy was found to associated with high risk of DSWI.

Fig. 2
figure 2

Open wound in the left anterior thorax (a) and chest CT (b). a The generator was removed and the RV lead remained. A subcutaneous pocket was formed in the midline and (b) extended to the anterior surface of the sternum (arrowheads)

Contrast-enhanced computed tomography (CT) showed that the superior vena cava (SVC) was completely occluded from the brachiocephalic vein to the SVC-RA junction by the four leads that had been implanted. The remaining two RV leads were detached to the middle of the SVC in the previous procedure. Based on these two findings, it was determined that extracorporeal circulation with RA-inferior vena cava (IVC) cannulation via the femoral vein would provide a bloodless field in the right heart system and lead extraction could be performed. Therefore, we decided to use an endoscopic minimal invasive cardiac surgery (MICS) approach to leads extraction.

A camera port was created on the right fourth intercostal mid-axillary line, a 3-cm main incision was made on the right fourth intercostal anterior axillary line, and a 5-mm left-hand port was made in the third intercostal space. Extracorporeal circulation was established by cannulation through the right femoral artery with an 18Fr. PCKC (SENKO MEDICAL INSTRUMENT Mfg.Co., LTD) and vein with a 25Fr. HLS cannula (Getinge). The wall of the right atrium was thickened. The RA was opened without cross-clamping after IVC snaring. The upper half of the RA lumen was occupied by an organizing thrombus and SVC was occluded. As there was backflow from the RV, repeated short circulatory arrests at 28℃ were used. Two ventricular leads were dissected and removed but one was heavily adherent to the RV and cut near the level of the annulus (Fig. 3). After completing the endocardial manipulation, a small incision was made in the left sixth intercostal space and an epicardial lead was placed in the left ventricular lateral wall. The generator was placed in the upper abdomen (Fig. 1c).

Fig. 3
figure 3

Intraoperative view. The right atrial wall was inflamed and thickened, and the upper half of the right atrium was occupied by an organizing thrombus. There were two right ventricular leads through the tricuspid annulus (arrowheads)

The clinical course was good and there was no surgical site infection. Tricuspid regurgitation was trace and didn’t change after surgery. The open wound was healed with radical debridement and segmental skin grafting after granulation (Fig. 4).

Fig. 4
figure 4

Healed wound. No surgical site infection was observed after device removal. Following granulation occurred, segmental skin grafting was performed and healing was achieved

Discussion and conclusions

The risk of a CIED infection has been reported to be 0.5%-1% and 1%-5% for a first-time procedure or replacement and upgrade, respectively. The associated mortality is not low (5%-17%) [1]. Treatment of a CIED infection requires the appropriate antibiotics and device removal. There are two main methods for CIED lead removal: TLE and surgical. TLE is a less invasive approach than surgery; however, the incidence of major complications is 1.8%-2.3% and catastrophic complications that require emergency surgical or endovascular intervention occur in 0.3%-0.8%. Moreover, the complication site is the SVC in 56% of cases [2, 3]. Lead-related risk factors for major complications include a greater number of extracted leads and a longer duration of lead implantation [3].

Surgical extraction is indicated in TLE failure cases, in patients at high-risk for complications, and in patients with complex lesions, large vegetations (> 2.5 cm), or the need for an epicardial lead implantation. The procedure generally requires a sternotomy and cardiopulmonary bypass, and therefore is more invasive than a TLE, and there is a high risk for DSWIs. Mediastinitis is associated with a mortality rate of 9.7%-29% and the long-term prognosis is poor [3,4,5,6]. The MICS approach may be less invasive than a surgical approach with a sternotomy and is more difficult to perform on peripheral veins than the SVC.

In our case, the wound was opened and the pocket was reached above the sternum. Surgical extraction was required due to TLE failure, but it was important to avoid making a median sternotomy. Fortunately, the remaining two RV leads were dissected from the SVC during the prior TLE procedure. Therefore, we could perform an endoscopic MICS approach via a small right thoracotomy. The MICS approach had three advantages. First, the RA wall was thickened by inflammation, therefore a lateral approach was appropriate for viewing the RV through the tricuspid valve. Second, the surgical incision was distant from the infected site in the anterior thorax bilaterally and a surgical site infection didn’t occur. Third, an injury for resternotomy could be avoided.

A hybrid TLE and MICS approach may be preferable in cases in which the TLE procedure alone is difficult or in patients at high risk for a DSWI.

Abbreviations

CIEDs:

Cardiac implantable electronic devices

CT:

Computed tomography

DSWI:

Deep sternal wound infection

IVC:

Inferior vena cave

MICS:

Minimal invasive cardiac surgery

MSSA:

Methicillin-sensitive Staphylococcus aureus

PM:

Pacemaker

RA:

Right atrium

RV:

Right ventricle

SVC:

Superior vena cave

TLE:

Transvenous lead extraction

References

  1. Cameron TL, Khaldoun GT. Cardiac implantable electronic device infection. Cleve Clin J Med. 2017;84(suppl 3):47–53.

    Google Scholar 

  2. Michael PB, Edmond MC, Oussama W, Bryan B, Walid IS, Joseph FS, et al. Outcomes of patients requiring emergent surgical or endovascular intervention for catastrophic complications during transvenous lead extraction. Heart Rhythm. 2014;11:419–25.

    Article  Google Scholar 

  3. Mitesh S, David TM, Tachel L, Jeptha PC, Craig P, Jude C. Incidence and predictors of perioperative complications with transvenous lead extractions: real-world experience with national cardiovascular data registry. Circ Arrhythm Electrophysiol. 2018;11:e004768.

    Article  Google Scholar 

  4. Kubota H, Miyata H, Motomura N, Ono M, Takamoto S, Harii K, et al. Deep sternal wound infection after cardiac surgery. J Cardiothorac Surg. 2013. https://doi.org/10.1186/1749-8090-8-132.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Milano CA, Kesler K, Archibald N, Sexton DJ, Jones RH. Mediastinitis after coronary artery bypass graft surgery Risk factors and long-term survival. Circulation. 1995;92:2245–51.

    Article  CAS  PubMed  Google Scholar 

  6. Braxton JH, Marrin CA, McGrath PD, Morton JR, Norotsky M, Charlesworth DC, et al. 10-year follow-up of patients with and without mediastinitis. Semin Thorac Cardiovasc Surg. 2004;16:70–6.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This study was not funded by any agency or institution.

Author information

Authors and Affiliations

Authors

Contributions

TO, and DT contributed to the success of this operation. SS, TS, HM, KT, II, and HU contributed to the manuscript drafting. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Shuji Nagatomi.

Ethics declarations

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nagatomi, S., Oshitomi, T., Teduka, D. et al. Removal of infected pacemaker leads using an endoscopic minimally invasive cardiac surgical approach: a case report. Gen Thorac Cardiovasc Surg Cases 3, 33 (2024). https://doi.org/10.1186/s44215-024-00156-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s44215-024-00156-4

Keywords